DIN Standard Medical Gas Terminal Unit (German Standard)
Reference Standards ISO 9170-1, YY 0801.1, DIN 13260
Functions and Performance Parameters
Applicable Gases: Oxygen ($O_2$), Vacuum (VAC), Medical Air (AIR), Nitrous Oxide ($N_2O$), Carbon Dioxide ($CO_2$), Instrument Air (IAIR), Driving Nitrogen ($N_2$).
Maximum Flow Rate: a) $> 200\text{ L/min}$ at a nominal pressure of $400\sim500\text{ kPa}$ ($O_2$, AIR, $N_2O$, $CO_2$) b) $> 350\text{ L/min}$ at a nominal pressure of $700\sim1000\text{ kPa}$ (IAIR, $N_2$) c) $> 25\text{ L/min}$ for Vacuum (VAC)
Maximum Withstand Pressure: $3\text{ MPa}$.
Plug-in/Unplug Cycle Life: $150,000\text{ cycles}$.
Service Life: $10\text{ years}$.
Inlet Connection Types: Copper pipe, flexible hose, threaded connection (M14, NIST, DISS), etc.
Equipped with a release/ejection protection mechanism to prevent accidental injury caused by probe pop-out.
Gas-specific pin-indexing (poka-yoke) design for the front and rear bodies to prevent hazards caused by gas cross-connection.
BS Standard Medical Gas Terminal Unit (British Standard) I. Reference Standards ISO 9170-1, YY 0801.1, BS 5682
Functions and Performance Parameters
Applicable Gases: Oxygen ($O_2$), Vacuum (VAC), Medical Air (AIR), Nitrous Oxide ($N_2O$), Carbon Dioxide ($CO_2$), Instrument Air (IAIR), Driving Nitrogen ($N_2$).
Maximum Flow Rate: a) $> 200\text{ L/min}$ at a nominal pressure of $400\sim500\text{ kPa}$ ($O_2$, AIR, $N_2O$, $CO_2$) b) $> 350\text{ L/min}$ at a nominal pressure of $700\sim1000\text{ kPa}$ (IAIR, $N_2$) c) $> 25\text{ L/min}$ for Vacuum (VAC)
Maximum Withstand Pressure: $3\text{ MPa}$.
Plug-in/Unplug Cycle Life: $150,000\text{ cycles}$.
Service Life: $10\text{ years}$.
Inlet Connection Types: Copper pipe, flexible hose, threaded connection (M14, NIST, DISS), etc.
Gas-specific pin-indexing (poka-yoke) design for the front and rear bodies to prevent hazards caused by gas cross-connection.
JIS Standard Medical Gas Terminal Unit (Japanese Standard) I. Reference Standard JIS T 7101
Functions and Performance Parameters
Applicable Gases: Oxygen ($O_2$), Vacuum (VAC), Medical Air (AIR), Nitrous Oxide ($N_2O$), Carbon Dioxide ($CO_2$).
Maximum Flow Rate: a) $> 200\text{ L/min}$ at a nominal pressure of $400\sim500\text{ kPa}$ ($O_2$, AIR, $N_2O$, $CO_2$) b) $> 40\text{ L/min}$ for Vacuum (VAC)
Maximum Withstand Pressure: $3\text{ MPa}$.
Plug-in/Unplug Cycle Life: $150,000\text{ cycles}$.
Service Life: $10\text{ years}$.
Inlet Connection Types: Copper pipe, flexible hose, threaded connection (M14, NIST, DISS), etc.
Gas-specific pin-indexing (poka-yoke) design for the front and rear bodies to prevent hazards caused by gas cross-connection.